The
Dielectric Spectroscopy Laboratory |
|
Low-Temperature
Dielectric Spectrometer
|
By
using a cell
mounted on
the cold head of a close cycle refrigerator, this
spectrometer allows to extend the temperature range down to 10 K
maintaining both a broad frequency range and a high
accuracy. (see
an
example below)
|
INSTRUMENT
CHARACTERISTICS
|
Impedance
Analyzer:
Solartron-Schlumberger SI1260 + Mestec dielectric module DM1370
-
Frequency
range: 1e-2
- 1e6 Hz
-
Sensitivity
better than
tan(delta) ~1e-4
Temperature
Control: Clod
head closed cycle He-refrigerator CTI-Cryogenics
-
Temperature
range: 10 -
320 K
-
Temperature
stability
better than ± 0.1 K
Samples:
Solids and viscous
liquids
-
A parallel plate capacitor of about 20 mm diameter
and
0.1 mm
thickens, formed with the sample by using two gold-plated
electrodes is located on the coldhead of the cryosystem.
|
|
|
|

|