| The
Dielectric Spectroscopy Laboratory | 
    
      | 
 | 
    
      | Low-Temperature
Dielectric Spectrometer
 
 | 
    
      | By
using a cell
mounted on
the cold head of a close cycle refrigerator, this
spectrometer allows to extend the temperature range down to 10 K
maintaining both a broad frequency range and a  high
accuracy. (see
an
example below) 
 
 | 
    
      | INSTRUMENT
CHARACTERISTICS | 
    
      | 
 Impedance
Analyzer:
Solartron-Schlumberger SI1260 + Mestec dielectric module DM1370
        
          Frequency
range: 1e-2
- 1e6 Hz 
          Sensitivity
better than
tan(delta) ~1e-4 
 Temperature
Control: Clod
head closed cycle He-refrigerator CTI-Cryogenics
        
          Temperature
range: 10 -
320 K 
          Temperature
stability
better than ± 0.1 K 
 Samples:
Solids and viscous
liquids
        
          A parallel plate capacitor of about 20 mm diameter
and
0.1 mm
thickens, formed with the sample by using two gold-plated
electrodes  is located on the coldhead of the cryosystem.
 
        
        
     
     
     | 
    
      |    
 | 
    
      |  
 
  
 |