The
Dielectric Spectroscopy Laboratory |
|
Micro-Wave
Dielectric Spectrometer |
|
This
spectrometer allows to access the microwave range over a limited
temperature range with a reasonable
accuracy. ( see
an
example below)
|
INSTRUMENT
CHARACTERISTICS
|
Analyzer:
Microwave network
analyzer Agilent
E8361A
-
Frequency
range: 2e7 -
7e10 Hz
-
Sensitivity
better than
tan(delta)
~1e-1
Dielectric
probe: Dielectric
probe kit Agilent
85070E
-
High
temperatue probe: temperature
range 250 -
470 K,
Frequency range 0.2-20
GHz
-
Performance
probe: temperature
range 250 -
470 K,
Frequency range 0.5-50
GHz
Temperature
Control:
Nitrogen-jet stream home-made system
-
Temperature
range: 100 -
570 K
-
Temperature
stability
better than ± 0.1 K
Samples:
Soft-solids and
Liquids
-
The sample is placed in a cylindrical aluminum
container
of 20 mm diameter, and the dielectric probe is placed on (or
inside) the sample.
|

|